MedicView hardware technology:
- Forward - reverse
PC - interface board communication
- PC & Windows
(2000, NT, XP) based
- 4-layer ISA gamma camera interface
card
- SME USB gamma camera interface
card
- BNC connections for X(positional),
Y(positional), E(energy), Z(logical) and G (gated)
signals
- X and Y analogue bipolar signals
in the range from +- 0.5 V to +- 10 V for the whole
gamma camera field of view
- Z (unblank) TTL level logical
signal with duration from 0.1microseconds
- G TTL level logical signal with
duration from 0.1 microseconds
MedicView software technology:
- Patient studies in
INTERFILE 3.3. standard
- Compatibility with other processing
software
- Software ZOOM settings: 1, 1.5,
2.0, and 2.5
- Software ORIENTATION settings
for image
- Matrix sizes: 64x64, 128x128
and 256x256 in word
- Acquisition modes: static, dynamic,
gated
- Stopping conditions: preset
time, preset counts and preset time/counts
- Software control of GAIN and
OFFSET setting signals
- On-line display of the images
with current time and count information during the
acquisition study
- Persistence scope function for
gamma camera tuning with regional count density evaluation,
changeable colour scales, size of ROIs, size of image
matrix, image smoothing, freezing, refreshing the scope
and selecting different display times.
|
|
|